1º Congreso Internacional de Medicina Crítica en Internet
CIMC´99

COMUNICACIONES
LENGUAS OFICIALES
LINGUAS OFICIAIS
OFFICIAL LANGUAGE

Español
English
Português

PANEL DE DISCUSIÓN
PANEL OF DISCUSSION
PAINEL DE DISCUSSÃO 


TEXTO COMPLETO

ABSTRACT
Nº 00110JA





 

Del 1 de Noviembre al 15 de Diciembre

ABSTRACT 
Nº 00110JA
TEXTO COMPLETO

PANEL DE DISCUSIÓN

DIA 10 A 17 DE NOVIEMBRE DE 1.999

Pacemaker with Autocaptura. Effects of body position on your perfomance.
TEXTO COMPLETO

Jose A. Lapuerta
* Oscar Sanz
Mª Teresa Antuña
Jose Guerra

Cabuenes Hospital. Gijon(Asturias) SPAIN

Introduction: 
Studies have show that Autocapture (St. Jude Medical , Pacesetter Inc.) system is reliable and saves pacing energy thanks to stimulation 0.3V above the threshold. Follow-ups are usually performed with the patient in supine position which perhaps is not the most common position during the day. This study attemps to determine differences in Autocapture-threshold, VARIO-threshold, Evoked Response detection and Spontaneous R-wave detection between supine and normal sitting position. 
From June 1997 to May 1999, 17 Microny SR+ pulse generator, model 2425T (St. Jude Medical, Pacesetter Inc.) and Membrane E, model 1450T (St. Jude Medical , Pacesetter Inc.) were implanted in 17 patients (5 female and 12 male) at our hospital. Indication in all of them was atrial fibrilation (AF) with severe AV block. Autocapture-threshold, Vario-threshold, Evoked Response (ER) and R wave (RW) were measured after One, Three, Six and 12 months in supine and sitting 90 degree. The pulse width used to perform Autocapture and Vario Threshold test was 0,3 ms (shipped setting).

RESULTS:
Data from 64 follow-ups have been collected:

                                   Supine(Mean ± SD)       Upright (Mean ± MD)      P value
VARIO (V)                                     0,74 ± 0,23                       0,75 ± 0,24               NS
AUTOCAPTURE (V)                     0,86 ± 0,24                        0,87 ± 0,23               NS
ER (mV)                                        9,18 ± 4,95                    10,14 ± 4,95          <0,001
RW (mV)                                      15,01 ± 7,32                     15,96 ± 8,10           0,001

CONCLUSIONS
Evoked responses values and spontaneous R-wave increased significantly when the patient is in an upright sitting position. The obseved changes did not interfere with the Autocapture algorithm and support routine  measurements in a supine position.
 

Topic=email session
Topic=Arritmias-Marcapasos
Topic=NO, TABLAS,FIGURAS
Name - EMail=jose.lapuerta@hcabuenes.es
 

TEXTO COMPLETO
u
 
This page was design by Dr. S. Herrero
© Critical Web Design, 1999